Performing Calibration (after verification failure)
Instrument calibration is required when the white verification measurement fails. The
verification measurement fails when the user defined white verification tolerance value is
exceeded. This tolerance is set on the Configuration Calibration dialog.
1.
Click
OK
on the “Out of White Verification Tolerance screen”.
2.
Make sure that the lenses on the VeriColor Spectro are clean (refer to the Cleaning
section later in this document) and then click
OK
on the “Insure lenses are …” screen.
3.
Attach the calibration tool as requested. Refer to Attaching the Calibration Tool as
previously explained. When the tool is properly attached, click the
OK
button on the
“attach calibration tube” screen.
4.
The instrument takes five calibration measurements and updates the Last Calibration
Time stamp.
5.
Click
OK
on the “Calibration Ok” screen.
6.
Remove the calibration tool and place in its storage bag.
NOTE:
If an error message appears after calibration, make sure the calibration tool is
located correctly over the illumination lens and is clean. You may also want to clean the
illumination lens and sensor lens on the VeriColor Spectro. Refer to the General Maintenance
section for cleaning procedures. After checking the calibration tool and cleaning the
instrument, click the
White Calibration
button again and repeat the procedure.
If a failure still occurs, contact X-Rite Customer Support at the number listed on the back
cover of this manual.