V E R I C O L O R
®
S P E C T R O
8
VeriColor Spectro Installation
Measurement Distance
Mount the VeriColor Spectro to the fixture using the holes provided at each flange.
VeriColor Spectro must be set to measure at a distance of 4
”
nominal to the measurement
surface. If the instrument is mounted in a 0°/30° geometry (preferred), the distance is that
from the center of the illumination lens to the sample. If the instrument is mounted in a
30°/0° geometry (alternate), the distance is measured from the center of the sensor lens to
the measurement surface.
CAUTION: Keep the sensor lens and illumination lens free from dust, smudge
marks, and finger prints.
4.0"
4.0"
Sample Part
Sensor Lens
Illumination Lens
Mounting
Flange (4)
0°/30° Mounting - Preferred
30°/0° Mounting - Alternate
Sample Part
Sensor Lens
Illumination Lens
Mounting
Flange (4)
IMPORTANT:
The Geometry setting in the
software will require editing from the default
setting of “0°/30°” if the alternate 30°/0° mounting
method is used. Refer to the Calibration
Configuration section later in this manual for the
procedure
.