Hardware Troubleshooting
This section deals primarily with troubleshooting problems with the PLC discrete inputs.
Problem Topic
Explanation/Solution
Terminal screws not tightened firmly on
wires.
Terminal not plugged in properly.
Loose or broken power wires.
Improper power connection.
Contamination has entered the terminal
and is preventing a reliable connection.
Over-current situation
caused by a short or
overload. Re-settable fuse
opens.
Fix the short or overload problem, wait a
little bit and re-apply power.
Over-voltage situation
(exceeds 30VDC). Re-
settable fuse opens.
Voltage spec is 24V +/- 10%. Correct
voltage supplied to VeriColor Spectro and
then re-apply power.
VeriColor Spectro will
not power up.
Power supply does not
provide enough current at
power-up.
Make sure power supply can provide at
least 3 Amps at start up.
“Read” input does not
work, system does
not measure.
LED indicator light does not
work (solid amber) with
input.
Instrument may have just powered up and
is still trying to initialize (flashing amber).
Wait until the VeriColor Spectro indicates a
solid “green” before attempting to measure.
Terminal screws not tightened firmly on
wires, or loose or broken wires.
Providing DC voltage or otherwise to the
input is not correct. Activating an input
entails pulling it low, usually with a ground
connection in the input terminal block.
Supplying too much voltage can cause
damage to the circuitry. Instrument will
need service.
Problems getting
discrete inputs to
work.
Most often the problem
would be wiring/electrical
problems. Each input has a
corresponding LED that will
light when the input is
active.
Inputs need to be held low for at least
100ms.
If the instrument is already busy with a
previous command, the input pulse may be
missed, as inputs are not buffered. It's OK
to hold down an input until an action is
observed. Another action won't be executed
until that input has been released.