V E R I C O L O R
®
S P E C T R O
51
•
Yellow light = One or more of the measurements displaying within the active trend
graph window has triggered the (warn condition) action limit. Yellow will continue to
display until the single measurement that caused the condition scrolls from the screen
in the trend graph (after 25 measurements), or a fail (red) condition occurs. In this
case, the light will change to red.
•
Red light = One or more of the measurements displaying within the active trend
graph window has failed. Red will continue to display until the single measurement
that caused the fail condition scrolls from the screen in the trend graph (after 25
measurements). At that point, the light could change to green or yellow if the current
measurement is in passing or warning condition.
Averaging (sliding average = 2 or more) and Only Last Record for PASS/FAIL enabled
(checked)
•
Green light = current grouped measurement average is a pass.
•
Yellow light = current grouped measurement average has triggered the (warn
condition) action limit.
•
Red light = current grouped measurement average is a fail.
Averaging (sliding average = 2 or more) and Only Last Record for PASS/FAIL disabled
(unchecked)
•
Green light = all of the grouped measurement averages displaying within the active
trend graph window are passing.
•
Yellow light = one or more of the grouped measurement averages displaying within
the active trend graph window has triggered the (warn condition) action limit. Yellow
will continue to display until the measurement group average that caused the
condition scrolls from the screen in the trend graph, or a failed (red) average
condition occurs. In this case, the light will change to red.
•
Red light = one or more of the grouped measurement averages displaying within the
active trend graph window has failed. Red will continue to display until the
measurement group average that caused the fail condition scrolls from the screen in
the trend graph. At that point, the light could change to green or yellow if the current
measurements average is in passing or a warning condition.
Additional Monitor Dialog Controls
-
Store
button is used to save the standard and sample measurement data in a .txt, .csv,
or .mif file format. For a .csv file format, refer to Importing a CSV File in the Color
Standards and Projects section for more detailed information on using this feature.
- Close
button closes the Monitor dialog returns you to the main application screen.
Summary of Contents for VeriColor Spectro
Page 1: ...VeriColor Spectro Non Contact Spectrophotometer Setup and Operator Manual ...
Page 2: ......
Page 25: ...V E R I C O L O R S P E C T R O 25 SAMPLE LOG REPORT ...
Page 65: ......
Page 66: ......
Page 67: ......