V E R I C O L O R
®
S P E C T R O
12
*Connect the “#5 – CH-D 8 / RS485-TX-A (gray)” and “#6 – CH-C 4 / RS485-RX-A (pink)” wires together for
Half Duplex. Also, connect the “#7 - CH-B 2 / RS485-RX-B (blue)” and “#8 – CH-A 1 / RS485-TX-B (red)”
wires together for Half Duplex.
**Use as common return to controller ground (insert a 10
Ω
resistor if voltage potential between the controller
and VeriColor Spectro is significant).
Power/Comm/Control Cable
(VCS50-EUR8-5DB)
#2 – Negative (brown)
#1 - +24 Vdc (white)
#5 – DSR (gray)
#6 – 10's (pink)
#7 – 1's (blue)
#8 – READ (red)
Cable Shield
(Panel or fixture ground)
Discrete Interface Cable
(SE108-EUR8-5M)
#1 – PASS (white)
#2 – FAIL (brown)
#3 – Output Comm (green)
#4 – CH-E 16 / 'ACTION' (yellow)
#5 – CH-D 8 / RS485-TX-A (gray)*
#6 – CH-C 4 / RS485-RX-A (pink)*
#7 – CH-B 2 / RS485-RX-B (blue)*
#8 – CH-A 1 / RS485-TX-B (red)*
Cable Shield**
DB9 Connector
Pin 5 – #2 Ground (brown)
Pin 2 – #3 Txout (green)
Pin 3 – #4 Rvin (yellow)
Summary of Contents for VeriColor Spectro