Service Manual
Troubleshooting and repair
21
Service menus for repair and network troubleshooting
In addition to the service menus supporting functional verification, these menus support
the following service procedures:
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Characterize NIBP
NIBP characterization sets the volume constants for the NIBP system to assure proper
cuff detection. You must characterize the NIBP system if any of the following occurs:
•
Characterization test failure
•
Pump failure
•
Internal tubing replacement (any or all)
•
Bleed valve, dump valve, or check valve replacement
•
Main board replacement or repair
1.
Attach a neonatal hose (008-0265-XX) to the monitor NIBP hose connector. Verify that
the hose length is 8 feet ± 4 inches (2.44 meters ± 10 cm).
2.
On the monitor, navigate to the
Characterize NIBP
screen.
HR/min
Setup
Service
NIBP
Characterize NIBP
.
Note
Before beginning any service procedure, read the entire procedure to verify that
you have all tools and equipment needed and that you understand each step.
Caution
Main board repair must be performed at the Welch Allyn Factory
Service Center. Any change to the main board requires system testing,
calibration, and verification, which can be done only at the Factory Service
Center.
Caution
Perform this characterization procedure only if you are prepared and
equipped to perform the
on page 61. To complete this
test, you must have the 35-ml test syringe described in
internal NIBP components.
internal NIBP components.
See the Propaq LT Service Manual
See the Propaq LT Service Manual
for the NIBP Characterization Procedure.
for the NIBP Characterization Procedure.
Show Info
Show Info
NIBP:
NIBP:
Pump Test
Pump Test
HALL, ROBERT E.
HALL, ROBERT E.
3456187
3456187
12:41:32P
12:41:32P
Adult,
Adult,
Rm 239
Rm 239
Audible Alarms Silenced
Audible Alarms Silenced
Service:
Service: NIBP
NIBP
ECG
ECG SpO2
SpO2 USB
USB Radio
Radio Software
Software
Display
Display Battery
Battery Voltages
Voltages Buttons
Buttons
Characterize NIBP
Characterize NIBP
Characterize NIBP after service to any
Characterize NIBP after service to any
Exit
Exit
Summary of Contents for Propaq LT 802LT0N
Page 1: ...Propaq LT Vital Signs Monitor Service Manual Software version 1 5X or later ...
Page 4: ...iv Contents Welch Allyn Propaq LT Vital Signs Monitor ...
Page 8: ...4 Safety summary Welch Allyn Propaq LT Vital Signs Monitor ...
Page 28: ...24 Troubleshooting and repair Welch Allyn Propaq LT Vital Signs Monitor ...
Page 74: ...70 Functional verification Welch Allyn Propaq LT Vital Signs Monitor ...
Page 78: ...74 35 ml test volume Welch Allyn Propaq LT Vital Signs Monitor ...