Service Manual
Troubleshooting and repair
17
Charger states and service
State Description
Charger
Status
LED
Charger
Power
LED
Type of State
Possible Corrective Action
0
Charger state
unknown.
n/a
n/a
Normal initial transitory state of
very short duration.
n/a
1
Monitor docked,
No-power fault.
Off
Off
Normal operating state when the
monitor is in the cradle and the
cradle is not powered.
If the power adapter is not
connected, connect it.
Fault state if the cradle fuse is
open.
Replace the fuse.
Yellow
On
Fault—power to the monitor
automatically disconnected by the
cradle. Possible monitor fault.
Attempt to recharge by removing
the monitor from the cradle and
reinserting the monitor in the
cradle.
If the fault persists, attempt to
recharge another monitor to
determine whether the problem is
with the cradle or with the
monitor main board.
Fault—no monitor in the cradle,
power input and input fuse OK.
Power board fault. Replace the
power board.
2
Monitor docked,
charging.
Green
Green
Normal operating state.
n/a
3
Monitor docked,
temperature
fault.
Yellow
Green
Fault—battery pack temperature is
out of range high or low.
Bring the battery pack
temperature into specified range
(0° C - 40° C).
4
Monitor docked,
battery voltage
too low for
charging.
Yellow
Green
Fault—battery pack can no longer
be charged.
Replace the battery pack.
“Removing the lithium ion battery
assembly”
5
Monitor docked,
timeout fault.
Yellow
Green
Fault—Battery pack failed to
charge within the specified time.
Attempt again to charge by
removing the monitor from the
cradle and reinserting the monitor
in the cradle.
Replace the battery pack.
“Removing the lithium ion battery
assembly”
6
Monitor docked,
charging
complete.
Off
Green
Normal operating state.
n/a
7
Monitor docked,
overcharge fault.
Yellow
Green
Fault—battery pack cells are
unbalanced or the cradle is faulty.
Replace the battery pack
(
“Removing the lithium ion battery
Replace the power board in the
cradle (
8
Monitor not
docked.
n/a
n/a
Normal operating state.
n/a
Summary of Contents for Propaq LT 802LT0N
Page 1: ...Propaq LT Vital Signs Monitor Service Manual Software version 1 5X or later ...
Page 4: ...iv Contents Welch Allyn Propaq LT Vital Signs Monitor ...
Page 8: ...4 Safety summary Welch Allyn Propaq LT Vital Signs Monitor ...
Page 28: ...24 Troubleshooting and repair Welch Allyn Propaq LT Vital Signs Monitor ...
Page 74: ...70 Functional verification Welch Allyn Propaq LT Vital Signs Monitor ...
Page 78: ...74 35 ml test volume Welch Allyn Propaq LT Vital Signs Monitor ...