4
25
Disassembly and reassembly
Procedures overview
This describes all disassembly steps required for module-level and board-level service
procedures.
Disassembly is presented here as a sequence of procedures.
Do not perform any unnecessary procedures.
Example: Remove the air filter assembly and replace the air filter
only
if it appears
that the filter is clogged and is restricting air flow to the pump.
Example: Disconnect and remove the NIBP pump
only
if you are going to replace the
pump or the NIBP air valve assembly.
Except where noted, reassembly is the reverse of the disassembly.
Reassembly notes are identified by underlined text.
Caution
Perform all repair procedures at a static-protected station.
Caution
Regard all monitor parts as extremely fragile, and execute all procedure
steps with care and precision.
Summary of Contents for Propaq LT 802LT0N
Page 1: ...Propaq LT Vital Signs Monitor Service Manual Software version 1 5X or later ...
Page 4: ...iv Contents Welch Allyn Propaq LT Vital Signs Monitor ...
Page 8: ...4 Safety summary Welch Allyn Propaq LT Vital Signs Monitor ...
Page 28: ...24 Troubleshooting and repair Welch Allyn Propaq LT Vital Signs Monitor ...
Page 74: ...70 Functional verification Welch Allyn Propaq LT Vital Signs Monitor ...
Page 78: ...74 35 ml test volume Welch Allyn Propaq LT Vital Signs Monitor ...