Service Manual
Overview
7
Product configurations
Model numbers for the configurations are as follows:
Recommended service intervals
Model
Monitoring Parameters
FlexNet/Acuity
Communication Enabled
Top-Level
Assembly
Serial #
Prefix
802LT0N
ECG, NIBP, Nellcor SpO
2
No
007-0160-XX
KA
802LTRN
ECG, NIBP, Nellcor SpO
2
Yes 802.11 FHSS
007-0161-XX
KA
802LTAN
ECG, NIBP, Nellcor SpO
2
Yes 802.11a
007-0384-XX
KA
802LT0S
ECG, NIBP, Masimo SpO
2
No
007-0268-XX
KA
802LTRS
ECG, NIBP, Masimo SpO
2
Yes 802.11 FHSS
007-0269-XX
KA
802LTAS
ECG, NIBP, Masimo SpO
2
Yes 802.11a
007-0383-XX
KA
Interval or condition
Action recommended
Procedure
Every 6 - 24 months
(per hospital protocols)
Verify functionality
Every 36 months
Replace the air filter
Every 36 months
Replace the battery
“Removing the lithium ion battery
assembly”
The battery does not hold a charge
Check battery capacity
Replace the battery
“Removing the lithium ion battery
assembly”
The monitor has been dropped or
otherwise damaged
Verify functionality
The monitor is not functioning
properly
Verify functionality
The monitor does not pass the
functional verification
Troubleshoot and repair, and then
verify functionality
Return the monitor to an authorized
Welch Allyn service center
Summary of Contents for Propaq LT 802LT0N
Page 1: ...Propaq LT Vital Signs Monitor Service Manual Software version 1 5X or later ...
Page 4: ...iv Contents Welch Allyn Propaq LT Vital Signs Monitor ...
Page 8: ...4 Safety summary Welch Allyn Propaq LT Vital Signs Monitor ...
Page 28: ...24 Troubleshooting and repair Welch Allyn Propaq LT Vital Signs Monitor ...
Page 74: ...70 Functional verification Welch Allyn Propaq LT Vital Signs Monitor ...
Page 78: ...74 35 ml test volume Welch Allyn Propaq LT Vital Signs Monitor ...