38
Disassembly and reassembly
Welch Allyn Propaq LT Vital Signs Monitor
REASSEMBLY: When reconnecting the tubing, lubricate the inlet port, using a cotton
swab dampened with water.
7.
Disconnect the ECG (flex) cable from ZIF connector J1 on the back of the main board.
Note
If the radio card or the antenna has been moved out of position, return the
monitor to a qualified Welch Allyn service center. The radio and the antenna are
not field-serviceable components.
antenna tucked inside the
chassis
foam tape (white) in place
ferrite bead in place
Air inlet filter port
NIBP tubing
Dampened cotton swab
ECG flex cable
J1