Service Manual
Overview
9
Related documents
Controls
Service menu
To access the monitor service menu:
From any main display screen (such as Large Numerics, Dual Waveform, ...):
1.
Highlight
,
HR/PR
,
SpO2
,
NIBP
,
Resp
, or
, and then press
.
2.
Highlight
Setup
(at the bottom of the screen) and press
. The setup menu appears.
3.
Scroll right ( ) to highlight
Service
, and press
. The service menu appears, with
NIBP
and
Show Info
highlighted.
Title
Reorder Number
Propaq LT Vital Signs Monitor Directions for Use
810-1828-XX
Monitor
Power - Turn the monitor power on or off.
Down - Move the cursor down to the next display
item.
Action - Act based on what is highlighted.
Silence/Reset - Silence the current alarm tone for
90 seconds or reset a silenced alarm tone.
Left - Move the cursor left to the next display
item; decrease the parameter value.
Display - Cycle to the next configured display
format, or cancel the current control, setup, or
pop-up menu.
Right - Move the cursor right to the next
display item; increase the parameter value.
Snapshot - Record a 21
-
second period of numeric
and waveform data.
Up - Move the cursor up to the next display
item.
Start/Stop NIBP - Start or stop an NIBP
measurement.
Cradle
Monitor Release - Press and hold while removing the monitor from the cradle.
II 1mV/cm
II 1mV/cm
Exit
Exit
Trends
Trends
Snapshots
Snapshots
Setup
Setup
Low
Low
Med
Med
80
80
HR/min
HR/min
140/78
140/78
NIBP mmHg (102)
NIBP mmHg (102)
Resp/min
Resp/min
SpO2
SpO2
Off
Off
On
On
Alarm Tone
Alarm Tone
Suspend Audible Alarms:
Suspend Audible Alarms:
Setup:
Setup:
(90 sec)
(90 sec)
12
12
97
97
%
%
HALL, ROBERT E.
HALL, ROBERT E.
3456187
3456187
3:00:06P
3:00:06P
Adult,
Adult,
Rm 239
Rm 239
Alarms
Alarms
ECG
ECG NIBP
NIBP Timings
Timings
High
High
Service
Service
Configuration
Configuration
Setup Menu
Summary of Contents for Propaq LT 802LT0N
Page 1: ...Propaq LT Vital Signs Monitor Service Manual Software version 1 5X or later ...
Page 4: ...iv Contents Welch Allyn Propaq LT Vital Signs Monitor ...
Page 8: ...4 Safety summary Welch Allyn Propaq LT Vital Signs Monitor ...
Page 28: ...24 Troubleshooting and repair Welch Allyn Propaq LT Vital Signs Monitor ...
Page 74: ...70 Functional verification Welch Allyn Propaq LT Vital Signs Monitor ...
Page 78: ...74 35 ml test volume Welch Allyn Propaq LT Vital Signs Monitor ...