Service Manual
Disassembly and reassembly
37
5.
Lift the top of the main board slightly away from the chassis to access the NIBP
tubing connection to the air inlet/filter.
6.
Disconnect the NIBP tubing from the air inlet/filter port.
REASSEMBLY: When you reassemble the bottom chassis, verify that the radio card and
the antenna are still in the correct positions:
Caution
Do not apply tool edges or serrations to the tubing. The tubing is
fragile, especially at the joints. To avoid damaging the tubing, disconnect using
only your fingers or unserrated pliers.
Main board screws
Reassembly: 4 inch lbs
Lift this side of the main board slightly
to access the NIBP tubing connection
underneath