Service Manual
Disassembly and reassembly
47
Disassembling the cradle
Replacing the fuse
3A, slow blow, 2AG
1.
Remove the fuse cover by rotating it (using a screwdriver) 90
counterclockwise.
2.
Remove and replace the fuse.
3.
Replace the fuse cover.
Removing the bed rail hook
1.
Remove the two screws securing the handle to the rear of the cradle.
2.
Slide the handle up and off of the shoulder screws.
Note
To facilitate reassembly, observe carefully while disassembling.
T3A/250V
12-28V
3A
fuse cover
Reassembly: 12 inch lbs