34 Pneumocheck Handle and Printer/Charger
Service Manual 610132 Rev. E
Each of the 3 calibration maneuvers was within 1/2% of
the average. The existing calibration factor will be
replaced with the cal factor from this calibration.
Shows calibration value after a successful calibration
was completed. The value is now the new calibration
value replacing the old one that was previously stored
in memory.
One or more of the calibration maneuvers exceeded
the average by more that 1/2%. The existing calibration
factor is not updated.
Test Values
FVC x.xxx
In Liters
FEV1 x.xxx
In Liters
RATIO xx
% FEV1/FVC
MMEF x.xxx
In liters per second
PEF x.xxx
In liters per second
MVV xxx
In liters per minute
RATE xxx
In liters per minute
TIME xx.x
In seconds
Appears when the power is recycled, indicating that the
memory is cleared.
Appears briefly when entering the test mode of the
device.
Indicates that the spirometer is transmitting an FVC test
to the printer base.
Indicates that the spirometer is transmitting an MVV
test to the printer base.
GOOD CAL
CF=XXX
RETRY.CAL
Status Messages
CLEAR
TEST.MODE
SEND FVC
SEND MVV
Section 3 - Theory of Operation
Summary of Contents for PneumoCheck 61000
Page 3: ...2 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Page 39: ...38 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Page 65: ...64 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Page 69: ...68 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ......