Service Manual 610132 Rev. E
Pneumocheck Handle Printer/Charger 37
The device cannot trim the baseline out. This is usually
due to an abrupt thermal change. The machine must
be allowed to return to an acceptable temperature
range before proceeding.
Encountering either the TEMP LIM or TRIM LIM
messages during normal operation may indicate a
possible hardware failure.
The spirometer failed to see a stable baseline value
within 30 seconds.
Bus error.
EEPROM check sum error. (This will occur once upon
first use of a new machine).
Out of memory.
Opcode error--another form of bus error.
Interrupt timing error.
EEPROM failure.
Spurious interrupt.
Error numbers greater that these indicate an
unexpected event in the program. In general, error
messages are probably caused by a hardware failure.
TRIM LIM
UNSTEADY
ERROR 0
ERROR 1
ERROR 2
ERROR 3
ERROR 4
ERROR 5
ERROR 6
Section 3 - Theory of Operation
Summary of Contents for PneumoCheck 61000
Page 3: ...2 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Page 39: ...38 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Page 65: ...64 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Page 69: ...68 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ......