Service Manual 610132 Rev. E
Pneumocheck Handle Printer/Charger 13
9
Depress & release test button: Ambient
temperature should be displayed.
20 degrees C = 68 degrees F
22 degrees C = 72 degrees F
24 degrees C = 75 degrees F
25 degrees C = 77 degrees F
27 degrees C = 81 degrees F
30 degrees C = 86 degrees F
If unit displays a reading outside ambient
temperature range, complete steps 10 thru 12.
Then, refer to troubleshooting guide. If unit
displays a reading of 390-400 degrees F,
check battery voltages. If battery voltage is
low, replace battery and repeat steps 4 thru
12.
10
Depress & release test button: Unit should
read FO = 55. This is a software coefficient.
11
Depress & release test button: Unit should
display a flashing ‘BAT OK’ message.
12
Depress & release test button: Unit should
display ‘DO DCV’ message.
13
Turn unit off. Depress test button and turn unit
on, release test button. Unit should read:
‘Test Mode’ >VER.3.5 >clear > select. The
latest software is VER 3.5: If unit reads VER
3.2, the software must be changed. The
PROM is part no 610102-535. It goes into
position U4 of the LCD digital PC board.
Section 2 - Service
Summary of Contents for PneumoCheck 61000
Page 3: ...2 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Page 39: ...38 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Page 65: ...64 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Page 69: ...68 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ......