22 Pneumocheck Handle and Printer/Charger
Service Manual 610132 Rev. E
A set of specification drawings have been included in
Section 7 with reference made to them here. The
following tests should be done before the unit is
returned to the customer.
1
Calibrate and test the handle per specification
drawing A01945. “Calibrating with Welch Allyn
Spirometer.
2.
Perform all tests on specification drawing
A01951, “Dynamic Waveform Testing”
For troubleshooting use, specification drawing
A01943 has been included for testing at the
PC Board Level.
1.
Perform Dielectric Strength Test per:
A00167- UL/CSA Models
A00178- IEC Models
2.
Perform Leakage Current Test per A00171
3.
Perform Grounding Continuity Test per:
A00172 - UL/CSA
A02242 - IEC
4.
Perform all tests listed on A01958, “Base Final
Testing”
For troubleshooting use, specification drawing
A01946 has been included for testing at the PC
Board Level.
2.4 TEST AND
ADJUSTMENTS
PNEUMOCHECK HANDLE
PNEUMOCHECK PRINTER/
CHARGER
Section 2 - Service
Summary of Contents for PneumoCheck 61000
Page 3: ...2 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Page 39: ...38 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Page 65: ...64 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Page 69: ...68 Pneumocheck Handle and Printer Charger Service Manual 610132 Rev E ...
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ......