DGS
5 Operation
Krautkrämer USM 100
Rev. - (09/2021)
163
Settings for the DGS measurement
– In the A-scan tap on the
icon to display
the function group.
– In the function
select
DGS
.
– Set all parameters in this function group (see from
page 138).
Recording a reference echo and turning the
DGS curve on
To be able to display the required DGS curve, you have
to record the reference echo.
– Peak the echo of the reference reflector, in this case
the backwall echo from the test object.
– After this, position the gate A on the reference echo
(see page 86).
(see page 66)
to record the reference echo.
The status icon
A DGS reference echo has been re-
corded
is displayed above the A-scan (see page 6).
– If not yet done, then set the function
to
On
in order to display the curve.
Summary of Contents for Krautkramer USM 10
Page 1: ...Krautkr mer USM 100 Operating Manual Id No 160M4395 Rev 09 2021...
Page 18: ...18 Rev 09 2021 Krautkr mer USM 100...
Page 19: ...Krautkr mer USM 100 Rev 09 2021 19 Introduction 1...
Page 34: ...1 Introduction How to use this manual 34 Rev 09 2021 Krautkr mer USM 100...
Page 35: ...Krautkr mer USM 100 Rev 09 2021 35 Standard package and accessories 2...
Page 43: ...Krautkr mer USM 100 Rev 09 2021 43 Initial start up 3...
Page 59: ...Krautkr mer USM 100 Rev 09 2021 59 Principles of operation 4...
Page 94: ...4 Principles of operation Multi color LED 94 Rev 09 2021 Krautkr mer USM 100...
Page 95: ...Krautkr mer USM 100 Rev 09 2021 95 Operation 5...
Page 177: ...Krautkr mer USM 100 Rev 09 2021 177 Documentation 6...
Page 191: ...Krautkr mer USM 100 Rev 09 2021 191 Maintenance and care 7...
Page 199: ...Krautkr mer USM 100 Rev 09 2021 199 Interfaces and Peripherals 8...
Page 205: ...Krautkr mer USM 100 Rev 09 2021 205 Appendix 9...
Page 219: ...Krautkr mer USM 100 Rev 09 2021 219 Specifications 10...
Page 225: ...Krautkr mer USM 100 Rev 09 2021 225 Index 11...
Page 241: ...11 Index Krautkr mer USM 100 Rev 09 2021 241...