5 Operation
DAC
158
Rev. - (09/2021)
Krautkrämer USM 100
Automatic variation of the measuring point in TOF
mode
An echo amplitude evaluation is normally carried out at
the echo peak of the signal under examination because
this is the only way of ensuring that the displayed echo
amplitude and the sound path (projection distance,
depth position) always belong to the highest echo in the
gate.
Note
The instrument checks the measuring point
setting in TOF mode before processing any
reference amplitudes. If
Peak
is not set as
measuring point, the instrument will automat-
ically switch to
Peak
. In this case, a note is
displayed at the bottom edge of the display
screen.
Summary of Contents for Krautkramer USM 10
Page 1: ...Krautkr mer USM 100 Operating Manual Id No 160M4395 Rev 09 2021...
Page 18: ...18 Rev 09 2021 Krautkr mer USM 100...
Page 19: ...Krautkr mer USM 100 Rev 09 2021 19 Introduction 1...
Page 34: ...1 Introduction How to use this manual 34 Rev 09 2021 Krautkr mer USM 100...
Page 35: ...Krautkr mer USM 100 Rev 09 2021 35 Standard package and accessories 2...
Page 43: ...Krautkr mer USM 100 Rev 09 2021 43 Initial start up 3...
Page 59: ...Krautkr mer USM 100 Rev 09 2021 59 Principles of operation 4...
Page 94: ...4 Principles of operation Multi color LED 94 Rev 09 2021 Krautkr mer USM 100...
Page 95: ...Krautkr mer USM 100 Rev 09 2021 95 Operation 5...
Page 177: ...Krautkr mer USM 100 Rev 09 2021 177 Documentation 6...
Page 191: ...Krautkr mer USM 100 Rev 09 2021 191 Maintenance and care 7...
Page 199: ...Krautkr mer USM 100 Rev 09 2021 199 Interfaces and Peripherals 8...
Page 205: ...Krautkr mer USM 100 Rev 09 2021 205 Appendix 9...
Page 219: ...Krautkr mer USM 100 Rev 09 2021 219 Specifications 10...
Page 225: ...Krautkr mer USM 100 Rev 09 2021 225 Index 11...
Page 241: ...11 Index Krautkr mer USM 100 Rev 09 2021 241...