Temperature Sensor Warnings
When excessive temperature is detected, the applicable icon displays in the
status bar, and one of the warning messages listed below appears across
the top of the display screen:
• Probe tip temperature has entered the WARNING Zone—this message
displays in an orange banner when tip temperature exceeds approximately
95°C and appears in the status bar.
• Probe tip temperature has entered the CRITICAL Zone—this message
displays in a red banner when tip temperature exceeds approximately
100°C and appears in the status bar.
• System is overheated, system shut down initiated—this message displays
in a red banner when internal temperatures exceed limits. Shutdown
automatically initiates and appears in the status bar.
When either the CRITICAL Zone or System Overheat warning appears,
take immediate action to lower the temperature to which the indicated
component is exposed.
Real3D™In Point Cloud Cursors
This feature allows users to adjust cursors while viewing an image in point
cloud. There are three options when adjusting cursors:
Inactive (represented as spheres)
Active (represented as small 3D ring)
Moveable (represented as large 3D ring)