![Waygate Technologies Everest Mentor Visual iQ VideoProbe Operating Manual Download Page 29](http://html1.mh-extra.com/html/waygate-technologies/everest-mentor-visual-iq-videoprobe/everest-mentor-visual-iq-videoprobe_operating-manual_3913187029.webp)
29
Press ‘next step’ softkey once complete.
Limit Line & Trigger Point
No adjustment of blade aerofoil is required within this step.
The Limit Line is displayed as a yellow line. Adjust cursors and position line to coincide with both Entry and Exit Lines (Figure 3a and 3b).
Fig 3a. Initial state post Exit Line assignment Fig 3b. Adjust Limit Line between both Entry and Exit Lines
The Trigger Point is rendered to the centre of the Limit Line. If required, adjust its location to the desired position. This precise point will coincide with the
Blade Counter incrementing one digit (up or down).
Assign ‘Ascend’ or ‘Descend’ Setting
It is possible to configure the Blade Counter to count either Ascending (factory default) or Descending. If the latter is required, press ‘Ascend’ softkey to
reveal ‘Decend’ softkey. The factory default is to count-up as the blade passes from the Entry to the Exit Line.
Note:
it is possible to make this adjustment later once the analytic is in operation via the Settings menu
Press
Next Step
softkey once complete.
Finalize Cursor Positions
Any cursors can now be checked and adjusted if necessary before completing the verification process.
For example, if probe was nudged and some or all points need to be amended (Figure 4).
Once complete press
Verify Blade Counter
softkey