Calibrating the 10 mA Reference Current
•
If the instrument’s 10 mA reference current is actually 10.262 mA, send
the following
Command Packet
to properly calibrate the unit.
Write
!010.26223
(
[10 mA BIAS CAL]
in mA)
Calibrating the 10 k
Ω
Resistance (for AD590)
•
If the instrument’s 10 k
Ω
resistance is actually 10.003 k
Ω
, send the
following
Command Packet
to properly calibrate the unit.
Write
!010.00327
(
[AD590 10 k
Ω
CAL]
in k
Ω
)
Reading the Sensor Bias Current and AD590 Sense
Resistance Settings
•
To read the sensor bias and AD590 sense resistance settings, the
Command Type changes from a WRITE (2) to a READ (1) and no
[PASSWORD]
is required:
Read
!000.00022
(Reads
[10
µ
A BIAS CAL] in
µ
A
)
!000.00021
(Reads
[100
µ
A BIAS CAL] in
µ
A
)
!000.00020
(Reads
[1 mA BIAS CAL] in mA
)
!000.00027
(Reads
[10 mA BIAS CAL] in mA
)
!000.00026
(Reads
[AD590 10 k
Ω
CAL] in k
Ω
)
Chapter 4 - Remote Interface Reference
Calibration - Sensor Bias Current and AD590 Sense Resistance
RS-232
Commands Discussed:
[10 mA BIAS CAL]
[AD590 10 k
Ω
CAL]
Summary of Contents for LFI-3751
Page 10: ...10 This page intentionally left blank ...
Page 13: ...1 Chapter 1 Quick Start 13 1 ...
Page 20: ...20 This page intentionally left blank ...
Page 21: ...2 Chapter 2 Front Panel Operation 21 2 ...
Page 34: ...34 This page intentionally left blank ...
Page 54: ...54 This page intentionally left blank ...
Page 68: ...68 This page intentionally left blank ...
Page 69: ...3 Chapter 3 Rear Panel Operation 69 3 ...
Page 75: ...3 4 Chapter 4 Remote Interface Reference 75 4 ...
Page 88: ...88 This page intentionally left blank ...
Page 127: ...5 Chapter 5 Specifications 127 5 ...
Page 131: ...Appendix CAT 220 Cable Accessory Diagram 131 ...