4
101
Chapter 4 - Remote Interface Reference
Enabling and Disabling the Output Current
Enabling and Disabling the Output Current
1. Enable Current
•
To enable the instrument’s output current send the following
Command
Packet
.
Write
!000.00123
(Enable Output Current)
2. Disable Current
•
To disable the instrument’s output current send the following
Command
Packet
.
Write
!000.00022
(Disable Output Current)
3. Monitor Output Status
•
To read the output status, the Command Type changes from a WRITE
(2) to a READ (1):
Read
!000.00021
•
The following table details the structure of the characters returned
in the data fi eld.
For complete detail about Enabling the Output Current, see Chapter 2.
Char 7 = Sign Char []
Char 6 = Not Used
Char 5 = Autotune Error Code
1 = Zero Value Current Limit Error
2 = Current Limit Cannot reach SET T
3 = Non-uniform TE I step measured
4 = Rate Sign Change
Char 4 = Autotune Status [0 = Normal Operation, 1 = Autotune Operation]
Char 3 = Decimal Point
Char 2 = Temp Limit or Sensor Error Status [1 = requires clearing]
Char 1 = Integrator Status [0 = OFF, 1 = ON]
Char 0 = Output Status [0 = OFF, 1 = ON]
RS-232
Commands Discussed:
[RUN / STOP]
If Character 2 is ‘1’, a Temp Limit or Sensor Error has occurred. The next
command to enable current will clear the error without enabling current.
Summary of Contents for LFI-3751
Page 10: ...10 This page intentionally left blank ...
Page 13: ...1 Chapter 1 Quick Start 13 1 ...
Page 20: ...20 This page intentionally left blank ...
Page 21: ...2 Chapter 2 Front Panel Operation 21 2 ...
Page 34: ...34 This page intentionally left blank ...
Page 54: ...54 This page intentionally left blank ...
Page 68: ...68 This page intentionally left blank ...
Page 69: ...3 Chapter 3 Rear Panel Operation 69 3 ...
Page 75: ...3 4 Chapter 4 Remote Interface Reference 75 4 ...
Page 88: ...88 This page intentionally left blank ...
Page 127: ...5 Chapter 5 Specifications 127 5 ...
Page 131: ...Appendix CAT 220 Cable Accessory Diagram 131 ...