4
95
Setting the Limit Currents
•
Separate positive and negative current limits are available:
[LIM I POSITIVE]
and
[LIM I NEGATIVE]
.
•
For example, to confi gure the instrument with maximum currents of
+1 Amp and -0.5 Amps, send the following two
Command Packets
.
Write
!001.00020
(Sets
[LIM I POSITIVE]
in Amps)
!101208-000.5002D
(Sets
[LIM I NEGATIVE]
in Amps)
•
To read the instrument’s limit current settings, the Command Type
changes from a WRITE (2) to a READ (1):
Read
!000.00022
(Reads
[LIM I POSITIVE]
in Amps)
!000.0002D
(Reads
[LIM I NEGATIVE]
in Amps)
•
[LIM I POSITIVE]
ranges from +000.000 to +005.000 Amps.
•
[LIM I NEGATIVE]
ranges from -005.000 to +000.000 Amps.
To Confi gure for a Resistive Heater
•
Set
[LIM I POSITIVE]
to +000.000 Amps to operate a resistive heater.
For complete detail about setting the Limit Current, see Chapter 2.
Chapter 4 - Remote Interface Reference
Setting the Limit Currents
RS-232
Commands Discussed:
[LIM I POSITIVE]
[LIM I NEGATIVE]
Summary of Contents for LFI-3751
Page 10: ...10 This page intentionally left blank ...
Page 13: ...1 Chapter 1 Quick Start 13 1 ...
Page 20: ...20 This page intentionally left blank ...
Page 21: ...2 Chapter 2 Front Panel Operation 21 2 ...
Page 34: ...34 This page intentionally left blank ...
Page 54: ...54 This page intentionally left blank ...
Page 68: ...68 This page intentionally left blank ...
Page 69: ...3 Chapter 3 Rear Panel Operation 69 3 ...
Page 75: ...3 4 Chapter 4 Remote Interface Reference 75 4 ...
Page 88: ...88 This page intentionally left blank ...
Page 127: ...5 Chapter 5 Specifications 127 5 ...
Page 131: ...Appendix CAT 220 Cable Accessory Diagram 131 ...