74
1. AC Power Entry Module
The
AC Power Entry Module
on the rear panel is designed to accept input
voltages from 98 to 133 and 196 to 265 VAC. Ensure that the power source
you plug the temperature controller into is properly grounded.
SELECT:
115V
(5x20mm SLO BLO)
FUSE (250V):
2000 maT @ 115V
1000 maT @ 230V
115/230 VAC, 50/60 Hz, 160 VA
LINE:
AC Power
Chapter 3 - Rear Panel Operation
AC Power
3. AC Fuse
For 115 VAC operation, install a 2000 maT fuse 5 x 20 mm SLO BLO fuse.
For 230 VAC operation, install a 1000 maT fuse 5 x 20 mm SLO BLO fuse.
2. AC Voltage Select Switch
The
Voltage Select Switch
confi gures the instrument for 115 or 230 VAC
operation. Before connecting power to your instrument, check the
Voltage
Select Switch
for proper AC voltage in your country. Only use the power
cord supplied with your instrument.
Do not change the setting of this switch while power is supplied to the
instrument. See the table below for the proper AC Voltage selection.
AC Voltage Available
98 -133 VAC
196 -265 VAC
Voltage Select
115 V
230 V
4. Earth Ground Connection
If your AC power system does not bring earth ground through the
AC
Power Entry Module
, then connect earth ground via a 16 AWG wire to
this screw terminal.
Summary of Contents for LFI-3751
Page 10: ...10 This page intentionally left blank ...
Page 13: ...1 Chapter 1 Quick Start 13 1 ...
Page 20: ...20 This page intentionally left blank ...
Page 21: ...2 Chapter 2 Front Panel Operation 21 2 ...
Page 34: ...34 This page intentionally left blank ...
Page 54: ...54 This page intentionally left blank ...
Page 68: ...68 This page intentionally left blank ...
Page 69: ...3 Chapter 3 Rear Panel Operation 69 3 ...
Page 75: ...3 4 Chapter 4 Remote Interface Reference 75 4 ...
Page 88: ...88 This page intentionally left blank ...
Page 127: ...5 Chapter 5 Specifications 127 5 ...
Page 131: ...Appendix CAT 220 Cable Accessory Diagram 131 ...