![Wavelength Electronics LFI-3751 User Manual Download Page 86](http://html1.mh-extra.com/html/wavelength-electronics/lfi-3751/lfi-3751_user-manual_3839339086.webp)
86
Chapter 4 - Remote Interface Reference
Diagnostic Command Code Listing
Command
Command
Type
Command
Code
Description and Data Field Range
LAMP CONTROL
PUSHBUTTON
SWITCH STATUS
ROTARY
SWITCH STATUS
DIGIPOT STATUS
2
1
1
1
Turns Front Panel Lamps on and off.
Char 7: Sign Char [+ = manual, - = automatic]
Char 6: Sequence Lamps [0 = OFF, 1 = Lamp ON]
Char 5: All Lamps ON [0 = OFF, 1 = Lamp ON]
Char 4: 7 Segment Displays [0 = OFF, 1 = Lamp ON]
Char 3: Decimal Point
Char 2: Unit of Measure & Shift LEDs [0 = OFF, 1 = Lamp ON]
Char 1: Status LEDs [0 = OFF, 1 = Lamp ON]
Char 0: Output ON LED [0 = OFF, 1 = Lamp ON]
OUTPUT, SHIFT, & TEMP LIMIT Pushbutton Status
Char 7 = Sign Char (Not Used); Returns ‘+’
Char 6, 5, 4 = (Not Used); Returns ‘000’
Char 3 = Decimal Point
Char 2 = Temp Limit Pushbutton [0 = OFF, 1 = ON]
Char 1 = Shift Pushbutton [0 = OFF, 1 = ON]
Char 0 = Output Pushbutton [0 = OFF, 1 = ON]
Rotary Switch Status
A ‘1’ indicates position of rotary switch
Char 7 = Sign Char (Not Used); Returns ‘+’
Char 6 = AUX T / C
Char 5 = LIM I / B
Char 4 = TE V / A
Char 3 = Decimal Point
Char 2 = TE I / D
Char 1 = SET T / I
Char 0 = ACT T / P
Read Digi-Pot counter
Char 7 = Sign Char: ‘+’ = clockwise, ‘-’ = counterclockwise
Char 6, 5, 4 = Integer Number of Interrupts since last cleared
Char 3 = Decimal Point
Char 2 = 0
Char 1 = 0
Char 0 = 0
91
92
93
94
Diagnostic Command Code Listing
When DIGI-POT is not moving, this command clears
the DIGI-POT counter. Once the DIGI-POT is
moving, this command reads the DIGI-POT counter.
The counter will contain the number of interrupts
generated since counter was last cleared.
DIGI-POT has 32 interrupts per rotation.
To check the lamps,
enter ‘+’ in Char 7 with
settings in Char 6 - 0.
To return to Normal
Operation, enter ‘-’ in
Char 7 or turn off
power.
Summary of Contents for LFI-3751
Page 10: ...10 This page intentionally left blank ...
Page 13: ...1 Chapter 1 Quick Start 13 1 ...
Page 20: ...20 This page intentionally left blank ...
Page 21: ...2 Chapter 2 Front Panel Operation 21 2 ...
Page 34: ...34 This page intentionally left blank ...
Page 54: ...54 This page intentionally left blank ...
Page 68: ...68 This page intentionally left blank ...
Page 69: ...3 Chapter 3 Rear Panel Operation 69 3 ...
Page 75: ...3 4 Chapter 4 Remote Interface Reference 75 4 ...
Page 88: ...88 This page intentionally left blank ...
Page 127: ...5 Chapter 5 Specifications 127 5 ...
Page 131: ...Appendix CAT 220 Cable Accessory Diagram 131 ...