23
2
Autotune PID Operation
Chapter 2 - Front Panel Operation
Autotune PID Operation
The LFI-3751 is capable of automatically characterizing your thermoelectric
or resistive heater load and calculating optimum control loop settings
for
P
,
I
, and
D
.
1. Autotune Process
During an Autotune, the LFI-3751 measures the load’s thermal lag time,
L
,
the maximum Temperature versus Time slope,
R
max
, and the load’s thermal
time constant,
T
, for a given output step current,
TE I
.
The time to complete an Autotune Optimization varies from load to load.
The process occurs in three phases:
Phase 1:
The LFI-3751 ramps the thermoelectric current to drive the thermal
load temperature toward the setpoint temperature.
Phase 2:
The LFI-3751 performs three characterization passes.
Phase 3:
The LFI-3751 reverts to normal operation, saving the optimized
P, I, and D settings in non-volatile memory and settling in at the
setpoint temperature.
When started,
Autotune generates
a step response
to characterize
your system.
TE Current
Time
Your system responds by
going towards the setpoint.
Temperature
Time
R
max
L
T
Ambient
Phase 1
First
Pass
Setpoint
Phase 2
Phase 3
Second
Pass
Third
Pass
Summary of Contents for LFI-3751
Page 10: ...10 This page intentionally left blank ...
Page 13: ...1 Chapter 1 Quick Start 13 1 ...
Page 20: ...20 This page intentionally left blank ...
Page 21: ...2 Chapter 2 Front Panel Operation 21 2 ...
Page 34: ...34 This page intentionally left blank ...
Page 54: ...54 This page intentionally left blank ...
Page 68: ...68 This page intentionally left blank ...
Page 69: ...3 Chapter 3 Rear Panel Operation 69 3 ...
Page 75: ...3 4 Chapter 4 Remote Interface Reference 75 4 ...
Page 88: ...88 This page intentionally left blank ...
Page 127: ...5 Chapter 5 Specifications 127 5 ...
Page 131: ...Appendix CAT 220 Cable Accessory Diagram 131 ...