Chapter 4 - Remote Interface Reference
Calibration - Measurement Slopes & Offsets
Calibration - Measurement Slopes & Offsets
•
All measurements are calibrated at the factory. To change these
calibrated values, you must fi rst send the
[PASSWORD]
command.
You should not use these commands without direct supervision from
Wavelength Electronics Technical Support Group.
Write
!101254????????(FCS)
(
[PASSWORD]
)
Calibrating ACT T Slope & Offset
•
These commands are used to calculate the instrument’s ACT T
measurement.
Write
!001.00024
(
[ACT T SLOPE CAL]
)
!000.00024
(
[ACT T OFFSET CAL]
in Volts)
Calibrating TE I Slope & Offset
•
These commands are used to calculate the instrument’s TE I
measurement.
Write
!001.00026
(
[TE I SLOPE CAL]
)
!000.00026
(
[TE I OFFSET CAL]
in Amps)
Calibrating TE V Slope & Offset
•
These commands are used to calculate the instrument’s TE V
measurement.
Write
!001.00028
(
[TE V SLOPE CAL]
)
!000.00028
(
[TE V OFFSET CAL]
in Volts)
RS-232
Commands Discussed:
[PASSWORD]
[ACT T SLOPE CAL]
[ACT T OFFSET CAL]
[TE I SLOPE CAL]
[TE I OFFSET CAL]
[TE V SLOPE CAL]
[TE V OFFSET CAL]
Summary of Contents for LFI-3751
Page 10: ...10 This page intentionally left blank ...
Page 13: ...1 Chapter 1 Quick Start 13 1 ...
Page 20: ...20 This page intentionally left blank ...
Page 21: ...2 Chapter 2 Front Panel Operation 21 2 ...
Page 34: ...34 This page intentionally left blank ...
Page 54: ...54 This page intentionally left blank ...
Page 68: ...68 This page intentionally left blank ...
Page 69: ...3 Chapter 3 Rear Panel Operation 69 3 ...
Page 75: ...3 4 Chapter 4 Remote Interface Reference 75 4 ...
Page 88: ...88 This page intentionally left blank ...
Page 127: ...5 Chapter 5 Specifications 127 5 ...
Page 131: ...Appendix CAT 220 Cable Accessory Diagram 131 ...