Chapter 4 - Remote Interface Reference
RESTORE Calibration Defaults
RESTORE Calibration Defaults
•
To issue this command you must fi rst send the
[PASSWORD]
command.
You should not use this command without direct supervision from
Wavelength Electronics Technical Support Group.
To revert this
instrument back to the default settings, send the following
Command
Packets.
Write
!101254????????(FCS)
(
[PASSWORD]
)
!000.00026
(
[RESTORE FACTORY
CALIBRATION
DEFAULTS]
)
•
NOTICE!
This command will clear the calibration E
2
PROM and clear
any previous calibration values. All slopes will reset to ‘1’ and offsets
to ‘0’. The sensor bias currents will become 10
µ
A, 100
µ
A, 1 mA,
and 10 mA. The AD590 sense resistance returns to 10 k
Ω
. The serial
number will revert to CALLWE.
RS-232
Commands Discussed:
[PASSWORD]
[RESTORE FACTORY
CALIBRATION DEFAULTS]
Summary of Contents for LFI-3751
Page 10: ...10 This page intentionally left blank ...
Page 13: ...1 Chapter 1 Quick Start 13 1 ...
Page 20: ...20 This page intentionally left blank ...
Page 21: ...2 Chapter 2 Front Panel Operation 21 2 ...
Page 34: ...34 This page intentionally left blank ...
Page 54: ...54 This page intentionally left blank ...
Page 68: ...68 This page intentionally left blank ...
Page 69: ...3 Chapter 3 Rear Panel Operation 69 3 ...
Page 75: ...3 4 Chapter 4 Remote Interface Reference 75 4 ...
Page 88: ...88 This page intentionally left blank ...
Page 127: ...5 Chapter 5 Specifications 127 5 ...
Page 131: ...Appendix CAT 220 Cable Accessory Diagram 131 ...