100
•
For example, after the sensor is properly confi gured, send the following
Command Packet
to confi gure the instrument with a resistance setpoint
of 10 k
Ω
.
Write
!010.00023
(Sets
[SET R]
in k
Ω
)
•
To read the instrument’s setpoint resistance, the Command Type changes
from a WRITE (2) to a READ (1):
Read
!000.00021
(Reads
[SET R]
in k
Ω
)
•
Resistance Setpoint ranges from +000.000 to +499.900 k
Ω
Chapter 4 - Remote Interface Reference
Setting Resistance Setpoint
Setting Resistance Setpoint
•
To use resistance as the setpoint, you must be using a resistive sensor:
thermistor or RTD. The following table shows the confi guration settings
and
Command Packets
to set the confi guration for a thermistor or RTD
in resistance mode.
How to Confi gure Thermistor and RTD for Resistance Setpoint
Sensor Type
Thermistor
RTD
Confi guration /
Setting
[A1] = 0
[A2] = 0
[A1] = 1
[A2] = 1
[B1] = 0
[B2] = 0
Command Packet:
Write
!000.00025
!000.00026
!001.00024
!001.00027
!000.00027
!000.00020
RS-232
Commands Discussed:
[A1]
[A2]
[B1]
[B2]
[SET R]
Summary of Contents for LFI-3751
Page 10: ...10 This page intentionally left blank ...
Page 13: ...1 Chapter 1 Quick Start 13 1 ...
Page 20: ...20 This page intentionally left blank ...
Page 21: ...2 Chapter 2 Front Panel Operation 21 2 ...
Page 34: ...34 This page intentionally left blank ...
Page 54: ...54 This page intentionally left blank ...
Page 68: ...68 This page intentionally left blank ...
Page 69: ...3 Chapter 3 Rear Panel Operation 69 3 ...
Page 75: ...3 4 Chapter 4 Remote Interface Reference 75 4 ...
Page 88: ...88 This page intentionally left blank ...
Page 127: ...5 Chapter 5 Specifications 127 5 ...
Page 131: ...Appendix CAT 220 Cable Accessory Diagram 131 ...