Chapter 4 - Remote Interface Reference
Calibration - Sensor Bias Current and AD590 Sense Resistance
Calibration - Sensor Bias Current and AD590
Sense Resistance
110
•
All sensor bias currents and the AD590 sense resistance are calibrated
at the factory. To change these calibrated values, you must fi rst send
the
[PASSWORD]
command. You should not use these commands
without direct supervision from Wavelength Electronics Technical Support
Group.
Write
!101254????????(FCS)
(
[PASSWORD]
)
Calibrating the 10
µ
A Reference Current
•
If the instrument’s 10
µ
A reference current is actually 9.89
µ
A, send the
following
Command Packet
to properly calibrate the unit.
Write
!009.89029
(
[10
µ
A BIAS CAL]
in
µ
A
)
Calibrating the 100
µ
A Reference Current
•
If the instrument’s 100
µ
A reference current is actually 99.5
µ
A, send the
following
Command Packet
to properly calibrate the unit.
Write
!099.50027
(
[100
µ
A BIAS CAL]
in
µ
A
)
Calibrating the 1 mA Reference Current
•
If the instrument’s 1 mA reference current is actually 1.005 mA, send the
following
Command Packet
to properly calibrate the unit.
Write
!001.00527
(
[1 mA BIAS CAL]
in mA)
RS-232
Commands Discussed:
[PASSWORD]
[10
µ
A BIAS CAL]
[100
µ
A BIAS CAL]
[1 mA BIAS CAL]
Summary of Contents for LFI-3751
Page 10: ...10 This page intentionally left blank ...
Page 13: ...1 Chapter 1 Quick Start 13 1 ...
Page 20: ...20 This page intentionally left blank ...
Page 21: ...2 Chapter 2 Front Panel Operation 21 2 ...
Page 34: ...34 This page intentionally left blank ...
Page 54: ...54 This page intentionally left blank ...
Page 68: ...68 This page intentionally left blank ...
Page 69: ...3 Chapter 3 Rear Panel Operation 69 3 ...
Page 75: ...3 4 Chapter 4 Remote Interface Reference 75 4 ...
Page 88: ...88 This page intentionally left blank ...
Page 127: ...5 Chapter 5 Specifications 127 5 ...
Page 131: ...Appendix CAT 220 Cable Accessory Diagram 131 ...