Chapter 2 - Front Panel Operation
Manually Setting the P, I, and D Constants
Manually Setting the P, I, and D Constants
Selecting optimum PID control parameters tha minimize overshoot and
maximize stability is even easier now when using the LFI-3751 with Autotune
PID. Refer to Autotune PID operation on page 23 for a detailed description of
the Autotune feature. The LFI-3751 can be confi gured to operate as a
P, PI,
PD
, or
PID
temperature controller.
The following sections discuss how to manually adjust the
P
,
I
, and
D
constants.
•
For most small loads, set the Proportional Gain to 30, the Integral Time
Constant to 1 second, and Differentiator Time Constant to OFF (0).
•
The
P, I,
and
D
settings are saved in non-volatile memory.
•
When fi rst controlling a load, adjust only the
P
term while leaving the
I
term at 1 second and the
D
term OFF. If the load will not stabilize for any
P
term, then increase the
I
term to 2 seconds. Again, adjust the
P
term
to optimize the system. Repeat these steps until the load temperature
stabilizes around the setpoint temperature. Use the
D
term to reduce
overshoot and to damp the system.
P
ACT T
LIM I
AUX T
TE V
SET T
TE I
B
C
I
A
D
decrease
increase
1. Adjusting the P Term
•
Rotate the
Display Select Switch
to
P
(Shift Button ON)
.
•
Rotate the
Digi-Pot Adjust Knob
to change the displayed value.
•
The Proportional Gain range is AutS, Autd, and 0 to 100.
•
RS-232 Command:
[P]
•
Increasing the proportional gain will improve overall temperature stability.
Since most thermal systems are “gain limited”, too high a proportional
gain can cause thermal cycling or instability in the instrument’s control
loop. If you notice the temperature of your load cycling around the setpoint
temperature, then reduce the proportional gain.
Display Proportional Gain
Default P, I, & D Constants
= 30
= 1 second
= 0 (OFF)
P
I
D
[P]
[I]
[D]
P
ACT T
LIM I
AUX T
TE V
SET T
TE I
B
C
I
A
D
Auto Setpoint
Auto Disturbance
OFF
ON
Proportional Gain Range
AutS
Autd
0.0
0.1 to 100
Summary of Contents for LFI-3751
Page 10: ...10 This page intentionally left blank ...
Page 13: ...1 Chapter 1 Quick Start 13 1 ...
Page 20: ...20 This page intentionally left blank ...
Page 21: ...2 Chapter 2 Front Panel Operation 21 2 ...
Page 34: ...34 This page intentionally left blank ...
Page 54: ...54 This page intentionally left blank ...
Page 68: ...68 This page intentionally left blank ...
Page 69: ...3 Chapter 3 Rear Panel Operation 69 3 ...
Page 75: ...3 4 Chapter 4 Remote Interface Reference 75 4 ...
Page 88: ...88 This page intentionally left blank ...
Page 127: ...5 Chapter 5 Specifications 127 5 ...
Page 131: ...Appendix CAT 220 Cable Accessory Diagram 131 ...