CLEANING THE CHAMBER FILTERS
Proceed as follows:
1
Allow the sterilization chamber to cool down.
2
Empty the sterilizer chamber by removing the trays and the rack.
3
Remove the filters at the bottom of the chamber (bottom/center) by unscrewing
them.
4
Rinse the filters with tap water.
5
Insert the filters in their original position, screwing them into place.
CLEANING THE EXTERNAL SURFACES OF THE STERILIZER
Proceed as follows:
1
Clean all external sterilizer covers with a slightly damp cloth moistened with
water.
Notice: Never use disinfectants, detergents or abrasive products.