Technical data
86
MN-111 | Instructions for use | MN-111 W&H Med ENG Rev11 | 2019/01/25 | © 2019 W&H Sterilization Srl
STANDARD STERILIZATION PROGRAMS
The sterilizer offers six preset FDA cleared sterilization programs that comply with the American National Standard ANSI/AAMI ST55:2010:
Program
Ref. intended
use
Recommended use
Sterilization
Temperature
Load example
1
Wrapped
Medium Load
Pouches
Medium Load
Instruments and dental handpieces, up to 3 lbs (1.4 kg).
Pouches, wrapped, unwrapped and naked.
270 °F (132 °C)
40 pouched instruments (probes, explorers,
mirrors, etc.),
4 pouched dental handpieces.
Wrapped Large
Load
Pouches Large
Load
Instruments and dental handpieces, up to 6 lbs (2.7 kg).
Pouches, wrapped, unwrapped and naked.
270 °F (132 °C)
80 pouched instruments (probes, explorers,
mirrors, etc.),
8 pouched dental handpieces.
Wrapped XL
Load
Wrapped
Cassettes
Instruments and dental handpieces, up to 14 lbs (6.4 kg).
Pouches, wrapped, unwrapped and naked.
270 °F (132 °C)
4 wrapped Hygiene cassettes (8” x 5.5” |
20.3 mm x 13.8 mm),
2 wrapped Restorative cassettes (8” x 11” |
20.3 mm x 27.9 mm),
4 wrapped Exam cassettes (8” x 3” | 20.3
mm x 7.6 mm)
Unwrapped
Unwrapped
Instruments and dental handpieces, up to 18 lbs (8.2 kg).
Unwrapped, naked.
270 °F (132 °C)
80 unwrapped instruments (probes,
explorers, mirrors, etc.),
8 unwrapped dental handpieces.
Low
Temperature
Low
Temperature
Textiles, up to 4.4 lbs (2 kg), or instruments and dental handpieces requiring
low temperature, up to 5 lbs (2.3 kg).
Pouches, wrapped, unwrapped and naked.
250 °F (121 °C)
65 pouched instruments (probes, explorers,
mirrors, etc.),
6 pouched dental handpieces.
Note
1
: Mixture of instruments and/or packages within each program may vary. Not to exceed maximum load weight.