Diagnostic
CONTENTS
This section deals with the following subjects:
Errors
CHECKS AND ACTIONS
Notice: For any error not listed in this table, call technical service.
Code
Description and actions
Actions
0xx
Load cannot be considered sterile. See "End of a
sterilization cycle" on page 53
Repeat the cycle.
If the problem persists, call service.
Check if the mains switch or network circuit breaker is OFF.
Check if the mains cable is properly connected.
Switch the sterilizer OFF and ON.
Set date and time, then switch the sterilizer OFF and ON.
Ensure that the sterilizer fan is not blocked.
10x
See error "13x to 16x" on the next page.
Repeat the cycle.
If the problem persists, call service.