Diagnostic
MN-111 | Instructions for use | MN-111 W&H Med ENG Rev11 | 2019/01/25 | © 2019 W&H Sterilization Srl
79
Code
Description and actions
Actions
18x
Chamber filter clogged. Remove and clean the chamber
filter. See error "13x to 16x" on the previous page.
Repeat the cycle.
If the problem persists, call service.
Bacteriological filter clogged. Check and replace if
necessary.
2xx
Switch the sterilizer OFF and ON.
Repeat the cycle.
If the problem persists, call service.
Wait for the chamber to cool down. Reset the thermal
overload (see "Extraordinary maintenance" on page 71).
3xx
Check the door seal. Clean or replace it if necessary.
Repeat the cycle.
If the problem persists, call service.
Clean the chamber face side.
Check the load does not exceed the MAXIMUM WEIGHT
LIMITS.
4xx
Clean water error (bad quality, clean tank low level, high
consumption of water).
Repeat the cycle.
If the problem persists, call service.
Drain and/or refill the clean water tank.
Check the door seal. Clean or replace it if necessary.
5xx
Check if there are hurdles on the door locking area
(chamber rack, loads, objects, …).
Repeat the cycle.
If the problem persists, call service.
Check the door seal (wrong placed).
Check if the door can move freely without touching the
trays or the load when closing.
Switch the sterilizer OFF and ON.
990
The cycle has been aborted by the user.
Re-process the load.