CUSTOM STERILIZATION PROGRAMS
WARNING! The custom programs are not FDA cleared and it is the responsibility of the user to validate these programs.
The sterilizer provides the following custom programs. These programs are disabled by default.
Program
Sterilization Temperature
Sterilization Time
(settable)
Drying time
(settable)
Custom A
250 °F (121 °C)
30–60 minutes
30–60 minutes
Custom B
270 °F (132 °C )
4–30 minutes
3–60 minutes
Custom C
273 °F (134 °C)
4–30 minutes
3–60 minutes
STERILIZATION PROGRAM OPTIONS (220 V VERSION)
WARNING! The sterility of items processed unwrapped is compromised on exposure to non-sterile environments. Ensure that items are
dry when removed from the sterilizer. If the default drying time is not adequate for the load to be sterilized, additional drying time can
be added (see "Sterilization cycle management" on page 49). To load properly the sterilizer see "Prepare the sterilizer" on page 48.
Unwrapped
Wrapped Medium
Load
Wrapped Large
Load
Wrapped XL Load
Low Temperature
Sterilization temperature
270 °F (132 °C)
270 °F (132 °C)
270 °F (132 °C)
270 °F (132 °C)
250 °F (121 °C)
Sterilization program type according ANSI/AAMI
ST55:2010
Dynamic-air-
removal: SFPP
Dynamic-air-
removal: SFPP
Dynamic-air-
removal: SFPP
Dynamic-air-removal: SFPP
Gravity displacement
Sterilization program type according EN 13060
Class S
Class S
Class S
Class S
Class S
Sterilization time (minutes)
4
4
4
4
30' 30"