Sterilization cycles
FILLING THE CLEAN WATER TANK
If no water system is connected you have to fill the tank:
1
Switch the sterilizer ON and open the chamber door.
2
Fill the clean water tank with distilled or demineralized water
until the sterilizer makes a sound. See "Technical data" on
page 93 for the tank capacity.
INSERTING THE CHAMBER RACK INTO THE STERILIZER
CAUTION! Risk of burns. Before touching the chamber rack
or contents, ensure the sterilization chamber is not hot.
1
Open the chamber door and align the chamber rack at the
center/bottom of the chamber.
2
Push the chamber rack gently into position until it clicks
into place.
3
Insert cassettes horizontally or vertically, or insert trays.
See "Load maintenance and preparation" on page 46 for
load requirements .
4
Close the door.
5
Turn the sterilizer switch ON: after the inizialization the
homepage appears.
GENERAL RECOMMENDATIONS
Follow these recommendations to obtain the most from the drying:
n
Ensure the paper side of the sterilization bags faces up, and
that the space in-between bags is enough.
n
To enjoy the full benefit of short cycle times, always place the
load on the uppertray of the chamber rack and remove other
trays from the chamber.
Sterilization cycle management
RUN A STERILIZATION CYCLE
1
On the homepage tap the desired
cycle.