Technical data
Technical data
WATER SUPPLY SYSTEM (TO BE COMPLIANT WITH IEC61770)
Fitting
Backflow preventing device (IEC61770).
Temperature
max. 95 °F (35 °C)
Pressure
min. 29 psi - max. 124.7 psi (min. 2 bar – max. 8.6 bar)
Flow
min. 0.066 - max. 0.132 gal/min (min. 0.25 – max. 0.5 l/min)
POWER SUPPLY SYSTEM
Nominal voltage and Max.
current
200–240 V ac, 50/60 Hz, 10 A, single-phase
100–125 V ac, 50/60 Hz, 12 A, single-phase
Overvoltage category
II
Protection required
Suitable circuit breaker and a Ground Fault Circuit Interrupter (GFCI).
All protection devices must be certified according to applicable
standard. A grounded connection is essential.
Communication with other
devices
2 USB port - 1 LAN port (optional)
Features
Fully micro-processor controlled, process evaluation system
according to EN13060.
Programmable standby mode.
Max. heat output
3000 kJ/h
INSTALLATION REQUIREMENTS
Working temperature
From +41 °F to +104 °F (from +5 °C to +40 °C)
Working relative humidity
Max. RH 80% up to 88 °F (31 °C), linearly decreasing to 50% at 104
°F (40 °C)
Storage temperature / rel.
humidity
From -4 °F to +140 °F (from -20 °C to +60 °C)/0–90 % (with empty
tanks)
Max altitude
3000 m asl
Min. atmospheric pressure
8.7 psi (0.6 bar)
Overall dimensions
W: 19.3’’/H: 17.8’’/D: 24.25’’ (W: 490 mm/H: 452 mm/D: 616
mm)
Min. space required
W: 23.22’’/H: 18.5’’/D: 24.25’’ (W: 590 mm/H: 470 mm/D: 616
mm)
Weight empty
105.8 lbs (48 kg)
Max. weight (fully loaded)
146.6 lbs (66.5 kg)
Weight per support area
39.2 kN/m
2
Enviroment pollution
Degree 2
STERILIZER CHAMBER
Pressure safety valve
37.7 psig (2.6 bar)
Safety thermostats
356 °F (180 °C)
Total volume
~7.4 gal, Ø: 11’’/D: 18’’ ~28 l, Ø: 279.4 mm/D: 456.8 mm
Usable space
~ 4.9 gal / 18.5 l
HEPA filter
0.3 µm