Sterilization cycles
46
MN-111 | Instructions for use | MN-111 W&H Med ENG Rev11 | 2019/01/25 | © 2019 W&H Sterilization Srl
CONTENTS
This section deals with the following subjects:
Load maintenance and preparation
Sterilization cycle management
Load maintenance and preparation
WARNINGS
WARNING! Any residual of chemicals (like cleaning and
disinfection products), could affect the purity of the steam
and consequently the whole sterilization process.
If necessary, the load shall be cleaned and lubricated in
accordance with the instrument manufacturer’s
instructions.
Notice: Any residual of chemicals could seriously damage the
sterilizer. The manufacturer’s warranty is void in case of damage
caused by chemicals.
CLEANING THE INSTRUMENTS
Clean all instruments thoroughly prior to sterilization. If possible,
clean instruments immediately after use; always follow the
instrument manufacturer ‘s instructions. Remove all traces of
disinfectants and detergents. Rinse and dry carefully all
instruments.
The instruments and tubes must be carefully rinsed and dried prior
to sterilization.
CORRECT LOAD PLACEMENT
WARNING! Do not overload trays and the chamber. Adhere to
the maximum load weight limits (see "Sterilization
programs" on page 85).
Wrap items with porous wrapping materials to facilitate steam
penetration and drying (e.g. sterilization bags for autoclaves).
Always use the chamber rack to allow adequate steam
circulation.
Follow these requirements:
Load type
Placement
Hinged instruments (e.g.,
forceps, extraction pliers, etc.)
In open position
Tubes
Place tubes on a tray allowing the ends to remain
open. Do not bend tubes.