MN-111 | Instructions for use | MN-111 W&H Med ENG Rev11 | 2019/01/25 | © 2019 W&H Sterilization Srl
55
Sterilization cycles
4
Check the message. See
"Messages of a stopped
sterilization cycle" below.
5
Tap to view the cycle parameters.
See "View the cycle parameters" on
page 52.
6
Open the chamber door.
7
Reprocess the load if necessary.
CAUTION! Hot steam. Wait
the steam to dissipate
before opening the door.
Note: Water could be present in the
chamber when opening the door. To
prevent spilling place a towel below
the chamber door.
MESSAGES OF A STOPPED STERILIZATION CYCLE
Following are the messages:
n
Load not sterile: Do not use items on patients!
n
Drying interrupted: The load might be wet. Wet items are for
immediate use only!
Unloading
WARNINGS
CAUTION! Risk of burns. Before touching, ensure the
sterilization chamber is cold. Always use the tray holder.
Sterilization cycle report
WHERE CYCLE DATA ARE STORED
The sterilizer stores in memory the summarized reports of the last
200 cycles and the analytical reports of the last 100 cycles. All
reports can also be saved on the USB pen drive or in a specific
remote folder in the network if the sterilizer is connected to a LAN
(optional). A maximum of 10000 cycles are storable in the provided
USB.