Maintenance
MN-111 | Instructions for use | MN-111 W&H Med ENG Rev11 | 2019/01/25 | © 2019 W&H Sterilization Srl
71
Extraordinary maintenance
DRAINING THE USED AND CLEAN WATER TANK
If you left accidentally the tanks full for more than seven days or if you plan not to use the sterilizer for at least seven days, you have to drain
the tanks.
1
Open the sterilizer service door.
2
Put a container below the sterilizer (1.84 gal (7 l) minimum) and place the end of
the drain tube in it.
3
To drain the used water, insert the drain tube connector in the grey port.
4
To drain the clean water, insert the drain tube connector in the blue port.
5
When the water has been completely drained, press the release button to remove
the drain tube and close the service door.
Summary of Contents for MN-111
Page 1: ...MN 111 110 V 220 V Instructions for use MN 111 W H Med ENG Rev11 ...
Page 2: ......
Page 103: ......