Diagnostic
Troubleshooting
MANAGING ERRORS
If during a sterilization cycle an error occurs do the following:
1
Wait until the OPEN button appears.
2
At the end of the reset phase, you
may open the door. A pop will
appear requiring a confirmation.
3
Tap
to open the door.
CAUTION! Do not switch off the
sterilizer during the reset
phase: it takes some minutes
to reset the system and reach
safe conditions in the sterilizer
chamber.
Notice: Water could be present in the
chamber when opening the door:
prevent spilling (e.g., place a towel
below the chamber door).
VIEW AND SAVE THE ERROR LOG
1
On the homepage tap
>
>
: a list of the last errors appears.
2
Tap
to save the list in the USB
pen drive.
ERROR PAGE
During the sterilization cycle, the sterilizer is continuously
monitored by a control system. If an anomaly is detected, the cycle
is aborted automatically, and the sterilizer starts a reset phase.