Technical data
MN-111 | Instructions for use | MN-111 W&H Med ENG Rev11 | 2019/01/25 | © 2019 W&H Sterilization Srl
89
Unwrapped
Wrapped Medium
Load
Wrapped Large
Load
Wrapped XL Load
Low Temperature
Sterilization temperature
270 °F (132 °C)
270 °F (132 °C)
270 °F (132 °C)
270 °F (132 °C)
250 °F (121 °C)
Sterilization program type according ANSI/AAMI
ST55:2010
Dynamic-air-
removal: SFPP
Dynamic-air-
removal: SFPP
Dynamic-air-
removal: SFPP
Dynamic-air-removal: SFPP
Gravity displacement
Sterilization program type according EN 13060
Class S
Class S
Class S
Class S
Class S
Sterilization time (minutes)
4
4
4
4
30' 30"
Drying time (minutes)
Range: 8-99
Recommended: 8
Range: 25–99
Recommended:
25
Range: 30–99
Recommended:
30
Range: 40–99
Recommended: 40
Range: 30–99
Recommended: 30
Total program duration (min)
1
Full loaded including drying time
43
53
58
71
80
Load type
Pouched, wrapped instruments and
dental handpieces
No
Yes
Yes
Yes
Yes
Unwrapped instruments and dental
handpieces
Yes
Yes
Yes
Yes
Yes
Unwrapped cassettes
Yes
Yes
Yes
Yes
Yes
Textiles
No
No
No
No
Yes
Maximum load weight (excluding trays)
instruments: 18
lb (8.1 kg)
instruments: 3 lb
(1.4 kg)
instruments: 6 lb
(2.7 kg)
instruments: 14 lb (6.3 kg)
included cassettes
instruments: 5 lb (2.2 kg)
included trays
or
textile: 4.4 lb (2 kg)
Note
1
: The total cycle time may vary depending on the type of load (solid or porous), the load weight, and other factors.
Summary of Contents for MN-111
Page 1: ...MN 111 110 V 220 V Instructions for use MN 111 W H Med ENG Rev11 ...