Getting started
Installing the sterilizer
LOCATION REQUIREMENTS
Notice:
Do not place the sterilizer so that it is difficult to operate the
controls behind the service door. Do not place the sterilizer so that
it is difficult to disconnect the power cord.
Leave the condenser grid (rear side of the sterilizer) free from
anything that might obstruct the air passage.
Surface materials should be water resistant. If sterilization cycles
will be continuous, pay attention to the surrounding materials:
steam can damage them.
The sterilizer must operate in absence of explosive atmospheres.
The sterilizer must operate in a well ventilated room far from
sources of heat and from flammable materials.
Place the sterilizer on a flat and level surface.
Clearance requirements to ensure proper air circulation:
ELECTRICAL CONNECTIONS
All the cables and tubes connected on the rear side of the sterilizer
must be placed far from the condenser grid (e.g. using the available
guides).
Notice:
Use only the cord set provided by the manufacturer.
Ensure that external and internal surfaces are free from moisture or
condensation before connecting to power.
The installation of the sterilizer shall be performed by two
authorized technicians using PPE (Personal Protective Equipment)
according to applicable standards.
The electrical power supply of the sterilizer must fulfill all applicable
standards in the country of use, and must comply with the data
label on the back of the sterilizer.