Maintenance
72
MN-111 | Instructions for use | MN-111 W&H Med ENG Rev11 | 2019/01/25 | © 2019 W&H Sterilization Srl
RESET THE SAFETY THERMOSTAT
The sterilizer is fitted with a safety thermostat to prevent it from overheating. If the safety thermostat operates because of too high
temperatures, the error 240 or a timeout error is displayed. The thermostat must be reset manually. Proceed as follows:
1
Wait for the sterilizer to finish the error reset phase and follow the instructions on
the display.
2
Switch the sterilizer OFF and remove the mains cable.
3
Wait for the sterilizer to cool down.
4
Open the sterilizer door.
5
Unscrew the cap and push on the reset button of the thermostat switch.
6
A click sound indicates that the thermostat switch has been reset.
7
Screw the cap.
8
Connect the mains cable and switch the sterilizer ON.
Note: If the thermostat operates repeatedly, call technical service.