Maintenance
68
MN-111 | Instructions for use | MN-111 W&H Med ENG Rev11 | 2019/01/25 | © 2019 W&H Sterilization Srl
1200-cycle or yearly maintenance
REPLACING THE DOOR SEAL
Notice: If you replace this consumable before the maintenance due date you have to reset the cycle counter. See "Replace the consumable
before the maintenance due date" on page 64.
Proceed as follows:
1
Open the chamber door.
2
Remove the used door seal by hand.
3
Carefully clean the seal seat and the inside face of the chamber door.
4
Moisten the new seal with water.