Maintenance
Note
2
: Even if the maximum cycle number is not reached, it is
recommended to replace the consumable parts every year, or if they
appear worn or damaged, or if the filters are clogged or discolored.
EXPIRED MAINTENANCE
The sterilizer monitors the wear of consumables by counting the
number of cycles executed since the last replacement.
When the number of cycles is close to the maximum, a pre-alert
about the concerned consumable is displayed. Please check that
you have the requested spare part available, buy one if not. When
the maximum number of cycles has been met, a message to
replace the consumable will be displayed.
1
Tap
to see an animated
replacement procedure.
2
When you have replaced the
consumable tap
to confirm: the
executed cycle counter is reset.
REPLACE THE CONSUMABLE BEFORE THE MAINTENANCE DUE
DATE
If you replace the consumables before the request of replacement
appears, you should manually reset the counters through the
following procedure.
1
On the homepage tap
>
2
Select the consumable you want to
replace: a message appears
showing the current worked hours
of the part.
3
Tap
to see an animated
replacement procedure.
4
When you have replaced the
consumable tap
to confirm: the
executed cycle counter is reset.