ONX-220 User's Guide
22135173, Rev. 000
June 2020
51
Chapter
2 Utilities
Updating the firmware from StrataSync
You can also connect to StrataSync via Ethernet to update the firmware of your unit.
1. Connect the OneExpert to the AC charger adapter to ensure an uninterrupted supply
of power during the update.
2. Establish a wired Ethernet connection from your instrument to the intranet or
network.
3. Verify the ONX has a valid IP address (it should have been changed from the default
address of 192.168.0.*)
4. Go back to the Home screen, scroll down to the bottom,
and select
StrataSync
.
5. On the
StrataSync
screen, enter the following:
y
StrataSync Account ID
– Determined at Setup
y
Interface
– Ethernet ; DOCSIS. If set to DOCSIS,
firmware upgrades will be skipped without
warning.
NOTE
: This setting does not select the
communication interface – Ethernet or RF/DOCSIS.
This setting must be made via the
CATV
screen
Connection
icon.