Swapping Devices
viewLinc 4.0 Administrator Guide
43
Note:
Each Vaisala Veriteq device must be reviewed
separately using vLog, and the COM port to
which they are connected must be selected.
•
COM port/IP address
.
Device is on the same COM port
(attached to the same cable). If you are swapping a 300
Series Transmitter connected as a network device,
ensure the new transmitter has the same IP address.
•
Name
.
Device is not retired or deactivated.
•
Status
.
Device is not in active alarm mode.
•
Sample Interval
.
Devices have the same sample
interval.
•
Channels
.
Devices have the same channels enabled
and be using the same units of measure for each
matching channel.
•
Audit Trail
.
Vaisala Veriteq devices are not linked to a
vLog Audit Trail.
2
From
Options | Locations
, navigate the system tree to
find the device you want to swap. Right click and select
Pause Device Alarming
.
3
In the Pause Device Alarming screen, you can add a
comment about why you are swapping devices, and
specify the expected duration for the pause. This will
ensure no communication or threshold alarms are
triggered when disconnecting the device.
4
Click
OK
.
5
Disconnect the device you want to swap, and connect
the new device. If you are swapping a 300 Series
Transmitter, unplug the power source and plug it in
again.
Summary of Contents for Veriteq viewLinc 4.0
Page 1: ...ADMINISTRATOR GUIDE viewLinc 4 0 Vaisala Veriteq viewLinc Software M211441EN A...
Page 10: ...About this Manual viii 2012 Vaisala Canada Inc...
Page 30: ...Chapter 1 Getting Started 20 2012 Vaisala Canada Inc...
Page 70: ...Chapter 3 System Settings 60 2012 Vaisala Canada Inc...
Page 86: ...Chapter 4 Locations 76 2012 Vaisala Canada Inc...
Page 108: ...Chapter 5 Alarms 98 2012 Vaisala Canada Inc...
Page 118: ...Chapter 6 Permissions 108 2012 Vaisala Canada Inc...
Page 124: ...Chapter 7 Events 114 2012 Vaisala Canada Inc...
Page 140: ...Chapter 9 Transfers 130 2012 Vaisala Canada Inc...
Page 146: ...Chapter 10 Mobile UI 136 2012 Vaisala Canada Inc...
Page 160: ...Appendix FAQs Troubleshooting 150 2012 Vaisala Canada Inc...
Page 168: ...Index 158 2012 Vaisala Canada Inc...